Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: AEMC 3930: (1 year) CAL VER /5500 DATE: 25-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 120 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required # Files: None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Press the following keys on the AMEC 3930 to 1.006 DISP to set up the UUT. 1.006 DISP 1.006 DISP (SHIFT) (CLR) 1.006 DISP (U eff) (> >) 1.006 DISP (RANGE U) (> >) 1.006 DISP (I eff) (> >) 1.006 DISP (P) (> >) 1.007 DISP Press the DISPL. key until the display reads: 1.007 DISP 1.007 DISP Ueff= 0 V 1.008 DISP Connect the 5500A and the UUT as follows: 1.008 DISP [32] 5500A NORMAL HI to UUT U* (+) Input 1.008 DISP [32] 5500A NORMAL LO to U (-) Input 1.009 HEAD {VOLTAGE} 1.010 DISP Select the 100 V range by pressing the RANGE U 1.010 DISP key until 100V is selected. Press the Ueff key 1.010 DISP to select Volts. 1.010 DISP 1.011 HEAD VOLTAGE : {100V Range} 1.012 5500 100.0V 0.3U 50H SI 2W 2.001 5500 10.0V 0.1U 50H SI 2W 3.001 DISP Select the 500 V range by pressing the RANGE U 3.001 DISP key until 500V is selected. Press the Ueff key 3.001 DISP to select Volts. 3.001 DISP 3.002 HEAD VOLTAGE : {500V Range} 3.003 5500 500.0V 1.5U 50H SI 2W 4.001 5500 50.0V 0.15U 50H SI 2W 5.001 HEAD {CURRENT} 5.002 DISP Connect 5500A to the UUT as follows: 5.002 DISP [32] 5500A AUX HI to UUT I* (+) Input 5.002 DISP [32] 5500A AUX LO to UUT I (-) Input 5.003 HEAD CURRENT: {1A Range} 5.004 DISP Select the 1A range by pressing the RANGE I 5.004 DISP key until 1A is selected. Press the Ieff key 5.004 DISP to select Amps. 5.005 5500 1000mA 4U 60H SI 2W 6.001 5500 100mA 1U 60H SI 2W 7.001 HEAD CURRENT: {5A Range} 7.002 DISP Select the 5A range by pressing the RANGE I 7.002 DISP key until 5A is selected. Press the Ieff key 7.002 DISP to select Amps. 7.003 5500 5.000A 0.02U 60H SI 2W 8.001 5500 0.500A 0.002U 60H SI 2W 9.001 HEAD CURRENT: {10A Range} 9.002 DISP Select the 10A range by pressing the RANGE I 9.002 DISP key until 10A is selected. Press the Ieff key 9.002 DISP to select Amps. 9.003 5500 10.00A 0.04U 60H SI 2W 10.001 5500 1.00A 0.01U 60H SI 2W 11.001 HEAD {POWER FACTOR} 11.002 DISP Select the 1A range by pressing the RANGE I 11.002 DISP key until 1A is selected. Select the 100V range 11.002 DISP by pressing the RANGE U until 100V is selected. 11.002 DISP 11.003 DISP Select the Power mode by pressing the P Key. 11.003 DISP the display should read: 11.003 DISP 11.003 DISP P-act.= 0.0W 11.004 HEAD POWER FACTOR: {100W Range (1A 100V)} 11.005 DISP Connect the 5500A to the UUT as follows: 11.005 DISP [32] 5500A NORMAL HI to UUT U* (+) Input 11.005 DISP [32] 5500A NORMAL LO to UUT U (-) Input 11.005 DISP [32] 5500A AUX HI to UUT I* (+) Input 11.005 DISP [32] 5500A AUX LO to UUT I (-) Input 11.006 M550 1A SI 11.007 5500 100V 60H SI S 2W 11.008 ACC 100W 0.0466U 11.009 MEMI Enter the UUT Power reading in Watts. 11.010 MEME 11.011 MEMC W 0.4U 12.001 M550 1A -90deg SI 12.002 5500 100V 60H SI S 2W 12.003 ACC 0.0W 0.2661U 12.004 MEMI Enter the UUT Power reading in Watts. 12.005 MEME 12.006 MEMC W 0.4U 13.001 END